Cleaning and Decontamination

Clean and low-level disinfect the Clean-Trace
Luminometer LX25 immediately if any spills occur or you suspect
contamination or after use in a room under contact precautions .
If any liquid is spilled inside the ATP Test chamber, do NOT
attempt to clean inside the test chamber itself . Contact your 3M
Health Care Service Center or local representative immediately .
Clean and disinfect the external surfaces of the Clean-Trace
Luminometer LX25 once per day at a minimum .
Cleaning and Disinfecting the External Surfaces:
1 . Power down the Clean-Trace Luminometer LX25 and
disconnect the USB/Power cable .
2 . Using a disposable towel lightly dampened with
disinfectant, gently wipe the outer surfaces of the
Clean-Trace Luminometer LX25 . Make sure the towel used
is just damp not dripping .
3 . When cleaning the Clean-Trace Luminometer LX25 avoid
the USB cable connection on the side of the Clean-Trace
Luminometer LX25 .
4 . Rinse disinfectant from Clean-Trace Luminometer LX25 by
repeating the procedure (2-3) using another disposable
towel slightly dampened with distilled or deionized water .
5 . Allow Clean-Trace Luminometer LX25 to dry for at least
1 hour .
6 . Do NOT reconnect the USB/Power cable until the
Clean-Trace Luminometer LX25 has been allowed to dry .
The following disinfectants may be used to clean and disinfect
the Clean-Trace Luminometer LX25 .
• Sodium Hypochlorite (chlorine bleach)
(up to .8% concentrations)
• Quaternary Ammonium Compounds (contains ammonium
chlorides, Isopropyl Alcohol, and/or Ethanol)

Contact Us

For questions or additional information on the
3M™ Clean-Trace ™ Luminometer LX25 call
the Clean-Trace Helpline at 1-800-228-3957 .