3M Unitek Ortholux Luminous Curing Ligh Kullanım Kılavuzu - Sayfa 15

Tıbbi Ekipman 3M Unitek Ortholux Luminous Curing Ligh için çevrimiçi göz atın veya pdf Kullanım Kılavuzu indirin. 3M Unitek Ortholux Luminous Curing Ligh 16 sayfaları. Fast curing cordless led light

3M Unitek Ortholux Luminous Curing Ligh Kullanım Kılavuzu
Guidance and manufacturer´s declaration – electromagnetic immunity
The Ortholux™ Luminous Curing Light unit is intended for use in the electromagnetic environment specified
below. The customer or the user of the Ortholux Luminous Curing Light unit should assure that it is used in such
an environment.
Immunity test
3 Vrms
Conducted RF
IEC 61000-4-6
150 kHz to 80 MHz
Conducted RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5 GHz
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in which the Ortholux Luminous
Curing Light unit is used exceeds the applicable RF compliance level above, the Ortholux Luminous Curing
Light unit should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the Ortholux Luminous Curing Light unit.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
IEC 60601 test
level
10 V
3 V/m
Compliance
level
Portable and mobile RF communications
equipment should be used no closer to any
part of the Ortholux Luminous Curing Light
unit, including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance
where p is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey,
level in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following symbol:
Electromagnetic environment -
guidance
80 MHz to 800 MHz
800 MHz to 2,5 GHz
b
a
should be less than the compliance
b