Absolute Process Instruments APD 6380 D Hızlı Başlangıç Kılavuzu

Verici Absolute Process Instruments APD 6380 D için çevrimiçi göz atın veya pdf Hızlı Başlangıç Kılavuzu indirin. Absolute Process Instruments APD 6380 D 4 sayfaları. Ac to dc transmitters, isolated, field rangeable

Absolute Process Instruments APD 6380 D Hızlı Başlangıç Kılavuzu
AC to DC Transmitters, Isolated, Field Rangeable
Input:
True RMS 0-40 mVAC to 0-300 VAC, 0-4 mAAC to 0-200 mAAC
Output:
0-1 V to 0-10 VDC, ±5 VDC, ±10 VDC, 0-2 mA to 20 mADC
O One Minute Setup for Hundreds of I/O Ranges
O External Switches & Tables for Range Selection
O Zero and Span Output Calibration Potentiometers
O Full 1200 V Input/Output/Power Isolation
O Input and Output LoopTracker
LEDs
®
O Output Test Button
O Built-In Loop Power Supply for Sink/Source Output
Applications
Q Convert, Boost, and Rescale Process Signals
Q Isolate Single-Ended (Common Ground) PLC Inputs
Q Interface Process Signals with Panel Meters, PLCs,
Recorders, Data Acquisition, DCS, & SCADA Systems
AC Input Ranges
Field selectable ranges via switch settings
Voltage:
0-40 mVAC to 0-300 VAC
Current:
0-4 mAAC to 0-200 mAAC
Measurements are true RMS
Input Impedance
Voltage (0-4 VAC): 1 MΩ minimum
Voltage (>4 VAC): 220 kΩ minimum
10 Ω typical
Current:
Input Frequency
40 Hz to 1000 Hz sinusoidal
Common Mode Rejection
120 dB minimum
LoopTracker
Variable brightness LEDs indicate I/O loop level and status
DC Output Ranges
Field selectable ranges via switch settings
Voltage:
0-1 VDC to 0-10 VDC, 10 mA max
Bipolar voltage:
±1 VDC to ±10 VDC
Current:
0-2 mADC to 0-20 mADC
20 V compliance, 1000 Ω at 20 mA
Output offset:
±100% in 15% increments
Output Calibration
Multi-turn zero and span potentiometers
±15% of span adjustment range typical
Output Loop Power Supply
20 VDC nominal, regulated, 25 mADC
Max. ripple, less than 10 mV
RMS
May be selectively wired for sinking or sourcing mA output
Output Test
Front momentary button sets output to test level
Potentiometer adjustable 0-100% of span
Output Ripple and Noise
Less than 10 mV
ripple and noise
RMS
Linearity
Better than ±0.1% of span
Ambient Temperature Range and Stability
–10°C to +60°C operating ambient
Better than ±0.02% of span per °C stability
Response Time
200 milliseconds nominal (0-90%)
Isolation
1200 V
minimum
RMS
Full isolation: power to input, power to output, input to output
Power
85-265 VAC, 50/60 Hz or 60-300 VDC, 2 W maximum
D versions: 9-30 VDC or 10-32 VAC 50/60 Hz, 2 W maximum
Housing and Connectors
IP 40, requires installation in panel or enclosure
For use in Pollution Degree 2 Environment
Mount vertically to a 35 mm DIN rail
Four 4-terminal removable connectors, 14 AWG max wire size
A
A
P
P
BSOLUTE
BSOLUTE
ROCESS
ROCESS
File E145968
85-265 VAC, 60-300 VDC
model only
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
Made in USA
Applications Link
api-usa.com/apps
Free Factory
I/O Setup!
1
Minute
Setup!
Pb
Lead
Free
Dimensions
0.89" W x 4.62" H x 4.81" D
22.5 mm W x 117 mm H x 122 mm D
Height includes connectors
Description
The APD 6380 accepts an AC voltage or current input and
provides an optically isolated DC voltage or current output that
is linearly related to the input.
Typical applications include signal isolation, conversion, boost-
ing or a combination of the three. Full 3-way isolation (input,
output, power) makes this module useful for ground loop
elimination, common mode signal rejection or noise pickup
reduction.
Sink/Source Output Versatility
For maximum versatility a milliamp output can be selectively
wired for sinking or sourcing. This allows the APD 6380 mil-
liamp output to connect to a powered or unpowered current
loop. The 20 VDC output loop supply can be used to power a
passive mA device if required.
How to Order
All models are field rangeable
For APD 6380, specify if UL version is required
Order options and accessories as required
Order D versions for operation on low voltage power
Model
Input
APD 6380
Field configurable—specify range if
factory is to set switches
APD 6380 D
Options–add to end of model number
M01
Output reversal, such as 20-4 mA output
U
Conformal coating for moisture resistance
I
I
NSTRUMENTS
NSTRUMENTS
Connect mA Output
1 2 3 4
for Sink or Source
Adjustable Output
5 6 7 8
Offset
Output LoopTracker
LED
Adjustable Output
Test Function
Zero and Span for
Output
Input LoopTracker
LED
Hundreds of Range
Selections
9 10 11 12
Universal
13 14 15 16
Power
LoopTracker
API exclusive features include two LoopTracker LEDs (green for
input, red for output) that vary in intensity with changes in the
process input and output signals. These provide a quick visual
picture of your process loop at all times and can greatly aid in
saving time during initial startup and/or troubleshooting.
Output Test
An API exclusive feature includes the test button to provide a
fixed output (independent of the input) when held depressed.
The test output level is potentiometer adjustable from 0 to
100% of output span.
The output test button greatly aids in saving time during initial
startup and/or troubleshooting.
Free factory setup. Please specify on your order
Input range
Output range
Output
Field configurable—specify range if
factory is to set switches
Accessory—order as separate line item
API BP4 Spare removable 4 terminal plug, black
1220 American Way Libertyville, IL 60048
Phone:
800-942-0315
Fax: 800-949-7502
APD 6380
Quick Link:
api-usa.com/6380
Removable Plugs
See Wiring
Diagrams on
Page 3
Power
85-265 VAC or 60-300 VDC
9-30 VDC or 10-32 VAC
api-usa.com