Emissions RF
Class A
CISPR 11
Table 2
Manufacturer's Declaration – Electromagnetic Immunity (IEC60601-1-2)
The Microflow-S is suitable for use in the specified electromagnetic environment. The customer and/or user
should assure that it is used in an electromagnetic environment as described below;
Immunity Test
Test Level
IEC 60601-1-2
Electrostatic
± 8 kV contact
discharge (ESD)
IEC 61000-4-2
± 2 kV ± 4 kV ± 8
kV ± 15kV air
Table 3
Manufacturer's Declaration - Electromagnetic immunity
The Microflow-S is suitable for use in the specified electromagnetic environment. The customer and/or user
should assure that it is used in an electromagnetic environment as described below;
Immunity
IEC 60601 test
Test
level
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 800
MHz
3 V/m
800 MHz to 2.5
GHz
Note 1: 80 MHz to 2.7GHz, the higher frequency range applies.
Note2 : These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radio, AM and FM radio broadcast, and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed
RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the Microflow-S is used exceeds the applicable RF compliance level
above, the Microflow-S should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the Microflow-S.
Over the frequency range 150 kHz to 80MHz, field strengths should be less than 3 V/m
6
Environmental data
6.1
Battery
Battery type: 9 V alkaline - 6LR61 or PP3 not rechargeable
Battery life: 5h
charge.
6.2
Physical specifications
Casing
Width = 80 mm
Depth = 150 mm
Height = 40 mm
Weight
200 g to 300g with battery and probe
6.3
Doppler sound output
Outpour power: 500mW RMS
20 174 C
Microflow-S pocket Doppler
User's Documentation
The emissions characteristics of Microflow-S make it suitable for use
in industrial areas and hospitals (CISPR 11 class A).
If it is used in a residential Microflow-S might not offer adequate
protection to radio-frequency communication services.
The user might need to take mitigation measures, such as relocating or
re-orienting this equipment.
Compliance
Level
± 8 kV contact
± 2 kV ± 4 kV ± 8
kV ± 15kV air
Compliance
Electromagnetic environment - guidance
level
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the
Microflow-S including cables, than the
3 V/m
recommended separation distance of 30cm or 12
inches.
Electromagnetic
Environment Guidance
Floor should be antistatic, wool. If floor are
covered with synthetic materials, the
relative humidity should be minimum at
least 35%.