6. PANEL LED INDICATION

6.1 [kg]: kg indicator turns on when weight in kg.
6.2 [lb]:
lb indicator turns on when weight in lb.
6.3 [stb]: stb indicator turns on when stable.
6.4 [zero]: zero indicator turns on when scale is in Zero position.
6.5 [tare]: tare INDICATOR turns on when this function is used.
6.6 [LoUW]: when sample unit weight is not heavy enough for accurate counting.
6.7 [LoWT]: when sample number is not enough for accurate counting.
6.8 [M+]:
is used when accumulating the counting data; this data is memorized
by pressing this key and can be accumulated up to 99 inputs.
6.9 [S1]: SCALE1 working state
6.10 [S2]: SCALE2 working state
6.11 [S3]: SCALE3 working state
6.12 [LoB]: low voltage, please charge battery.
6.13 [CHG]: charge status. Red LED turns on when charging. Red and green LED
lights will be flashing alternately after the battery is fully charged.
7.
CALIBRATION:
Calibration may be required if the scale is used in a different gravity location than
initially calibrated.
7.1 Scale 1 Calibration
Turn on the scale, then press <1>, <3>, <1>, <1>, <Enter> when the scale
examining itself.
7.1.1 Menu Setting:
'WEIGHT display' shows SCALE1,
'UNIT WEIGHT display' shows CAL-0.
'TOTAL display' will show A/D value.
Then press <UNIT> button to select calibration units between kg and lb.
7.1.2 Calibration:
Make sure nothing loaded on the bigger platform.