- Turn on scale
- As the scale is counting down, press the following buttons in order [1], [3], [1], [3]
then press ENTER
- The following must be completed befor;e beginning the calibration process for SCALE 3:
- WEIGHT DISPLAY shows SCALE 3
- UNIT WEIGHT shows "xxxx" (this is the full capacity of SCALE 3)
- TOTAL DISPLAY will show "d=xx" (this is the division of SCALE 3)
- Press UNIT key to select full capacity in kg or lb
- Press CE Key to set the full capacity – key in the number through the numeric keypad
- Press (*) to set the division – key in the number through the numeric keypad
- Press ENTER to confirm
- Now begin the CALIBRATION process:
- Press UNIT key to select whether you will calibrate in kg or lb
- Press ENTER to set zero point (UNIT WEIGHT will show CAL-0 flashing for a
few seconds and then changes to XXXX to set the calibration weight)
- Press [CE] key and then press the corresponding number buttons based off of the
amount of weight you are using to calibrate (example: if you are using 1lb to calibrate the
scale you would key in 1lb).
- Calibration weight will show in the UNIT WEIGHT display.
- Press the ENTER key, the UNIT WEIGHT display will show flashing XX and then
changes to XX.XXX lb. Once this occurs the calibration process is complete.
-
***Note: Make sure nothing is loaded on the platform before you are asked to place the
calibration weight. Once calibration is complete you will need to reboot the scale and place the
sample test weight on the platter to assure the calibration was completed correctly.
6 Maintenance and service
The weighing machine must be treated carefully and cleaned regularly. It is a precision instrument.
For maintenance-work, the balance must be separated from the power supply (remove power
adapter plug from socket). Also ensure that the balance cannot be reconnected to the power supply
during the work by a third party.