Anyload ES310 Kullanıcı Kılavuzu - Sayfa 8
Terazi Anyload ES310 için çevrimiçi göz atın veya pdf Kullanıcı Kılavuzu indirin. Anyload ES310 12 sayfaları. Postal & shipping scale
c. "ADD10" means that 10 objects to be weighed should be added onto the
platform. Attention: The weight of each sample should be higher than the
scale's minimum division value; otherwise, the window will display wrong cue
"no".
d. If you want to change the sampling quantity, please press the
button
to change the quantity; if you do not need to sample again, press the
button to cancel the sampling, and the scale will adopt last sampling value to
count ;
e. After you confirm the quantity of objects on the platform and the data
displayed on the LCD are the same, short press
to confirm, and then begin
to sample and count;
f. During the process of sampling and counting, the screen will display the
sampling data by rapidly flickering; keep the platform calm, stable, and without
other interference factors which will influence the weighing performance. If the
weighing stays stable longer than 2s, the sampling and counting will be
automatically finished, the parameter will be saved automatically and the scale
will return to counting mode; if not, the screen will keep flickering the data and
waiting for the stability of measuring.
6.9 Serial Port Printout
a. Connect the serial port with the scale according to the drawing below. Baud
Rate=9600,without parity digit, sent stop bit 1 (The serial connection diagram
is omitted);
PC or Printer
PS Scale
b. Open the serial communication software, for example, "HyperTerminal" in
windows. Select the connected portal number "COM" and set the baud rate as
9600bps, the parity bit as "NONE", the data bit as "8Bit", and the stop bit as
"1Bit".
c. You can select the serial output modes according to your own needs. The
output modes include 3 types: output by pressing
button, automatic
output after the weight keeps stable, and continuous output. You can set
though the "out" in "Parameter setting".
ANYLOAD ES310 Postal & Shipping Scale User's Manual (V1611)
7 |