Dansensor CheckPoint O2 Kullanıcı Kılavuzu - Sayfa 7
Ölçüm Cihazları Dansensor CheckPoint O2 için çevrimiçi göz atın veya pdf Kullanıcı Kılavuzu indirin. Dansensor CheckPoint O2 11 sayfaları.
12
User Guide
What does "Response time (T
According to the specifications on page 3 the O
sensor has a Response time (T
This means that when performing consecutive
measurements in areas with large differences in
the O
concentrations, the device will reach 95 %
2
of the "true" value during the first measurement
(15 sec.).
O reading
2
[%]
Start value
T value
95
"True" value
PBI-320297-B 02/2014
)" mean
95
) of 15 sec.
95
Difference
T=0
EN
Example:
(See illustration below):
2
• Last measurement was performed in i.e.
20.9 % O
• The following value to be measured is
1.0 % O
2
• Difference is:
20.9 - 1.0 = 19.9 % O
• The expected read-out value after first
measuring is:
20.9 - (0.95 * 19.9) = 2.0 % O
Difference = Start
T value = Start value
95
T=15
(1. measurement)
Consequently this means that if there is a large
difference in the O
consecutive measurements, multiple
measurements must be performed to obtain the
most accurate result.
COPYRIGHT ©
CheckPoint O
2
2
2
value
"True" value
(0.95*Difference)
T=30
(2. measurement)
concentrations between two
2
Dansensor A/S
2
time [sec]