Aplisens WW-11ALW Kullanıcı Kılavuzu - Sayfa 15

Monitör Aplisens WW-11ALW için çevrimiçi göz atın veya pdf Kullanıcı Kılavuzu indirin. Aplisens WW-11ALW 20 sayfaları.

Aplisens WW-11ALW Kullanıcı Kılavuzu
EN.IO.WW.11.ALW
9.

OPERATION

The LCD has three primary information fields in the figure belowa as LCD1, LCD2, LCD3.

Figure 8. Display information fields.

LCD1 - value of current or guidance percent preset range display. Depending on the display
configuration will be displayed the current value of the current line 4...20 mA with a resolution of
0.1 mA, or percent guidance the preset range with a resolution of indications 1%.
LCD2 – display field for the digital value measured by display, the value rescaled by user's units, and
error codes. The position of the decimal point can be set in the local MENU.
LCD3 – information field. During normal operation it is designed for continuous display of the base unit
or the user units. In the MENU operation mode it displays the setting options. It is also used to display
errors related to the execution of commands in the local setup MENU.
Display backlighting - local display is equipped in backlight which can be switched on or off depend-
ing on needs. Switching the backlight on increases the voltage drop for all versions by 3 V. Enable or
disable the backlight is possible using the jumper after removing the back cover of the display. The
method of switching the backlight on or off is shown in figures below.

Figure 9. View of disassembled display unit.

15
Revision 01.B.003/2020.09