Sensor Calibration
Apogee SQ-620 Extended Range PFD Sensors have standard calibration factors of exactly:
Multiply the calibration factor by the measured mV signal to convert sensor output to photon flux density in units
of µmol m
-2
s
-1
:
Calibration Factor (20 µmol m
20
LED Output
(1000 µmol m
20 µmol m
-2
-1
s
per mV) * Sensor Output Signal (mV) = Photon Flux Density (µmol m
*
-2
-1
s
)
Sensor Output
50 mV
-2
-1
s
per mV
50
=
1000
Example of photon flux density measurement with an
Apogee model SQ-620 Extended Range PFD Sensor. The
LED in this example Full yields a photon flux density on a
horizontal plane of 1000 µmol m
signal of 50 mV. This signal is converted to photon flux
density by multiplying by the calibration factor of 20 mol
-2
-1
m
s
/ mV.
-2
-1
s
)
-2
-1
s
. This yields an output