Apogee Instruments SIF-111-SS Kullanıcı El Kitabı - Sayfa 7
Ölçüm Cihazları Apogee Instruments SIF-111-SS için çevrimiçi göz atın veya pdf Kullanıcı El Kitabı indirin. Apogee Instruments SIF-111-SS 19 sayfaları. Infrared radiometer
Spectral Response
Spectral response of SIF series
infrared radiometers. Spectral
response (green line) is determined
by the germanium filter and
corresponds closely to the
atmospheric window of 8 to 14 µm,
minimizing interference from
atmospheric absorption/emission
bands (blue line) below 8 µm and
above 14 µm. Typical terrestrial
surfaces have temperatures that
yield maximum radiation emission
within the atmospheric window, as
shown by the blackbody curve for a
radiator at 28 C (red line).