APPA ATL-PV Talimatlar

Test Ekipmanları APPA ATL-PV için çevrimiçi göz atın veya pdf Talimatlar indirin. APPA ATL-PV 2 sayfaları. Test leads

ATL-PV Test Leads
Instruction
Probe tip guard cap
Finger Guard
For CAT III or CAT IV environments, use the test leads
with the probe tip guard cap fixed firmly. Without the
probe tip guard cap, the test leads can be used in
CAT II environment ONLY.
For 1500V AC & 2000 V DC measurement, This test
lead can only be used in the environment that is not
connected to MAINS directly.
 Caution
Make sure that test leads are firmly connected to the
V-COM terminals of the correct instrument, and the
instrument have to switch to PV mode.
 Warning
•When using test leads or probes, keep your fingers
behind the finger guards.
•Use caution with voltages above 30 Vac rms, 42 Vac
peak, or 60 Vdc. These voltages pose a shock
hazard.
•If the test lead is used in a manner not specified by
the manufacturer, the protection provided by the
equipment may be impaired.
•To reduce the risk of fire or electric shock, do not use
this product around explosive gas or in damp
locations.
•DO NOT USE the test leads when the internal white
insulation layer is exposed.
•DO NOT USE the test leads above maximum ratings
of CAT. environment, voltage and current, that are
indicated on the probe and the probe tip guard cap.
•DO NOT USE the test leads without the probe tip
guard cap in CAT III and CAT IV environments.
•DO NOT USE the test leads to measure over
1000V that is connected to MAINS directly.
Maintenance
Do not attempt to repair this test lead set. It contains no
user-serviceable parts. Repair or servicing should only
be performed by qualified personnel.
Cleaning
Clean the test lead with a water and mild detergent. DO
NOT use abrasives or solvents and DO NOT IMMERSE
in liquid.
Specification
Input Impedance: 10MΩ
Overvoltage Category: CAT II 1000V AC ,1500V DC
CAT III 1000V, CAT IV 600V.
Pollution Degree 2
Exposed probe tip length :
19 mm to 4 mm (0.75 inch to 0.16 inch)
Environmental ratings :
-10°C to 45°C (-4°F to 113°F), 80% R.H.
Altitude : 2000 m (6,562 ft)
Safety Standard : EN61010-031
CAT
Application field
The circuits directly connected to Low-voltage
installation.
The building installation.
The source of the Low-voltage installation.