DataApex Shimadzu LC-10 Manuel - Sayfa 17

Su Pompası DataApex Shimadzu LC-10 için çevrimiçi göz atın veya pdf Manuel indirin. DataApex Shimadzu LC-10 19 sayfaları. Metal free pump

Shimadzu LC-10/20 Pump

4.4 Shimadzu LC-10/20 Pump Setup

Shimadzu LC-10/20 Pump Setup
Configuration
type and set some other basic parameters.
Port
Sets the COM port to which the Shimadzu LC- 10/20 Pump pump is
connected.
Pump Model
Allows to select from several types of Shimadzu pumps. LC-10ADvp and
LC-10ATvp models are supported at the moment.
Note:
If any LC- 20 type pump should be connected, it is necessary to set the
LC-10 emulation mode on the pump (for more details see the chapter
Installation Procedure on pg 3)..
Valve Model
Allows to select the valve model that may be installed on the pump to
create low pressure gradient.
Sys Mode
Sets the pump operation mode. Isocratic and Low Pres. Grad. options are
available.
Note:
The pump hardware must be set accordingly (for more details see the
chapter Installation Procedure on pg 3)..
Solvents
Sets the number of solvents when the Low Pres. Grad. mode is used.
Max Pressure
Sets the maximum pressure allowance of the pump. This value is sent to
the pump and when the pressure exceeds this value, pump will react on it.
The pressure check is also present in Clarity, in the
Gradient - Gradient Options
dialog) is used to set the connection to the pump, select it's
Fig 10: Shimadzu LC-10/20 Pump Setup
dialog. Value of Max Pressure set there
4 Using the control module
dialog (accessible through the
- 13 -
System
Method Setup - LC