Decade 1021 Vac Back Chair
8
No Air from the Syringe
1. Is the Master On/Off toggle in the ON position?
Yes: Proceed with Step 2.
No: Move the toggle to the ON position
and retest.
2. Is the air compressor ON and operating
properly?
Yes: Proceed with Step 3.
No: Correct any problems in the compressor,
then retest the unit.
3. Is the manual air shutoff valve in the floor box
fully OPEN (turned fully counterclockwise)?
Yes: Proceed with Step 4
No: Turn the valve fully counterclockwise,
then retest the unit.
4. Is the syringe air flow pinch valve
completely closed?
Yes: Turn the air flow adjustment screw
counterclockwise to the desired flow,
then retest the unit.
No: Look for pinched or crimped tubing
between the floor box and the syringe.
No Water from the Syringe
1. Does the Indicator on the floor box turn
red when the Master On/Off Toggle is in
the ON position?
Yes: Proceed with Step 2 on page 9.
No: Make sure signal air is reaching the air-
actuated water shutoff valve. Look for
pinched or crimped signal air tubing.