Troubleshooting & Maintenance
Problem
BlueWave® MX-275
LED does not
produce light
BlueWave® MX-275
LED suddenly stops
producing light
BlueWave® MX-275
LED provides only
low-intensity light
Product Cleaning and Care
▪
Product cleaning is limited to wiping the product with a damp cloth. Do not soak.
Isopropanol Alcohol or household cleaners may be used for cleaning the product.
▪
Always inspect the quartz window for cleanliness before use. Foreign material can cause
permanent damage to the window. Clean with Isopropanol Alcohol to remove smudges or
foreign material. Damaged or permanently etched windows should be replaced.
▪
Do not use compressed air to removed particle debris inside the emitter as it may damage
the high-speed cooling fan.
Possible Cause
LED intensity adjustment
set to 0% or too low
LED cycle time is set to
0 seconds
Interlock is open
Interface cable connections
loose or damaged
Trigger setting not
matched to input.
LED head is not connected
to the correct port/channel
Over-temperature
shutdown was triggered
Footswitch defective
Interlock is open.
LED intensity adjustment
set to minimum
Contaminated/dirty lens
optics
BlueWave® MX-275 Emitter User Guide
Corrective Action
Increase LED intensity setting.
0 Seconds sets manual mode and
requires a trigger.
Verify interlock jumpers are in place.
Verify PLC command structure for PLC
mode.
Check connections and condition of
interface cable.
Trigger setting on admin screen should
match the desired input trigger
channel.
Verify that the head is connected to the
desired port/channel.
Verify alarms.
Activate unit using the front control
panel. Replace the footswitch if the unit
operates from the front control panel.
Verify interlock jumpers are in place.
Verify PLC command structure for PLC
mode.
Increase LED intensity setting on admin
settings or I/O input for PLC mode.
Clean the surface of the lens.