Gait up PhysiRun Lab Kullanıcı Kılavuzu - Sayfa 14

Analitik Cihazlar Gait up PhysiRun Lab için çevrimiçi göz atın veya pdf Kullanıcı Kılavuzu indirin. Gait up PhysiRun Lab 17 sayfaları.

Gait up PhysiRun Lab Kullanıcı Kılavuzu

5.2 Support

LED indications
Situation
Description
Connected
via
Green continuous
USB
Orange blinking
Orange blinking
Red continuous
During measure
Green blinking (2 flashes per blink)
Green blinking (2 flashes per blink)
Green/Orange blinking (1 green, 1 orange flash per
blink)
Error
2 rapid red blinks
3 rapid red blinks
4 rapid red blinks
5 rapid red blinks
6 rapid red blinks
Error type
Error type*
How to solve the issue
Error 1
Disconnect and replug the Physilog
Error 2
Contact GaitUp
Error 3
Charge the Physilog
Error4
Contact GaitUp
Error 5
Contact GaitUp
Error 6
Empty the memory card
*see LED indication above
Soft reset of Physilog
The soft reset is the first manipulation which can be done if the Physilog
properly. To do a soft reset, hold the main button for 15 seconds. The LED shines orange
and then stops. If the problem persists, fill out the contact form on the support page
(www.gaitup.com/support) or write an email to [email protected]. Please indicate
®
Physilog
5 serial number and describe the problem in as much detail as possible.
User guide PhysiRun Lab - v1.1.1 Date: 06.05.2019
Copyright © 2018-2019, All Rights Reserved - Gait Up SA. CH-1015 Switzerland
®
5
®
5
EPFL Innovation Park, Bat C. CH-1015 Lausanne
Meaning
Fully charged
Charging
Data transfer
Error 1
Measuring
Waiting for synchronization
Low battery
Error 2
Error 3
Error 4
Error 5
Error 6
®
5
Gait Up SA
www.gaitup.com
TVA CHE-145.136.760
®
5 is not working
14