iHealth BP3 Kullanım Kılavuzu - Sayfa 5

Kan Basıncı Monitörü iHealth BP3 için çevrimiçi göz atın veya pdf Kullanım Kılavuzu indirin. iHealth BP3 11 sayfaları. Fully automatic arm cuff electronic blood pressure dock
Ayrıca iHealth BP3 için: Kullanıcı Kılavuzu (12 sayfalar), Hızlı Başlangıç Kılavuzu (2 sayfalar)

iHealth BP3 Kullanım Kılavuzu
allow the blood circulation in your arm to be stabilized. Prolonged over inflation of
the bladder may cause ecchymoma of your arm.
6.User is not permitted to change the battery. If the battery is out of life, please
send to the service center for maintenance.
7.This Electronic Sphygmomanometer is designed for adults and should never be
used on infants or young children. Consult your physician or other health care
professionals before use on older children.
8.Blood pressure measurements determined by this Dock are equivalent to those
obtained by a trained observer using the cuff/stethoscope auscultation method,
within the limits prescribed by the American National Standard, Electronic or
automated Sphygmomanometer.
9.Please avoid strong magnetism interference, such as microwave oven, etc.
10.It is recommended to set the iPod touch, iPhone, or iPad in Airplane mode
during measuring to avoid strong magnetism interference. If a call comes in during
the measurement, the measurement process will be terminated automatically.
11.If Irregular Heartbeat (IHB) brought by common arrhythmias is detected in the
procedure of blood pressure measurement, a signal of
will be displayed.
Under this condition, the Electronic Sphygmomanometer can keep functioning,
but the results may not be accurate, it's suggested that you consult with your
physician for accurate assessment.
There are 2 conditions under which the signal of IHB will be displayed:
1) The coefficient of variation (CV) of pulse period >25%.
2) The difference of adjacent pulse period≥0.14s and the number of such pulse
takes more than 53 percentage of the total number of pulse.
12.Please do not use the cuff other than supplied by the manufacturer, otherwise it
may bring biocompatible hazard and might result in measurement error.
13.The Dock might not meet its performance specifications if stored or used
outside the specified temperature and humidity ranges in specifications.
14.Please do not share the cuff with other infective person to avoid cross-infection.
15.Medical AC adapter which output is DC 5.0V and complies with IEC 60601-1/UL
60601-1 and IEC 60601-1-2/EN 60601-1-2 is suitable for this Dock, such as
ASP12-05021002JU (input: 100-240V, 50/60Hz, 200mA; output: DC 5V, 2.1A). Please
note that the Dock jack size: USB Type B.
If the AC adapter is abnormal, please change the adapter.
Note: Do not pull out the adapter when you are using the Dock!
Do not use any other type of AC adapter as it may harm the Dock!
16.This device complies with part 15 of the FCC Rules. Operation is subject to the
following two conditions: (1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that
4