GENERAL SAFETY AND PRECAUTIONS

1. Read all of the information in the Owner's Manual and other provided
instructions before operating the unit.
2. Consult your physician for any of the following situations:
a)The application of the cuff over a wound or inflamed area.
b)The application of the cuff on any limb with intravascular access or therapy,
or an arterio-venous (A-V) shunt.
c)The application of the cuff on the wrist on the side of a mastectomy.
d)Simultaneous use with other medical monitoring equipment on the same
limb.
e)The blood circulation of the user needs to be checked.
3. Do not use this product in a moving vehicle as this may result in inaccurate
measurements.
4. Blood pressure measurements determined by this product are equivalent to
those obtained by professional healthcare practitioners using the
cuff/stethoscope auscultation method within the limits prescribed by the
American National Standard, Electronic or Automated Sphygmomanometer.
5. When a call comes in during the measurement, the measurement process
will be terminated automatically. It is recommended that the iOS device be
set in Airplane mode during measurement to avoid strong magnetism
interference.
6. If Irregular Heartbeat (IHB) is detected during the measurement procedure,
the IHB symbol will be displayed. Under this condition, the Wireless Blood
Pressure Wrist Monitor can keep functioning, but the results may be
inaccurate. Please consult your physician for accurate assessment.
There are 2 conditions under which the signal of IHB will
be displayed:
1)The coefficient of variation (CV) of pulse period >25%.
2)The difference of adjacent pulse period ≥0.14s and the number of such
pulse takes more than 53 percent of the total number of pulses.
7. Please do not use any other cuff other than that supplied by the
manufacturer as this may result in measurement errors and a biocompatible
hazard.
8. Information regarding potential electromagnetic or other interference
between the blood pressure monitor and other devices together with advice
regarding avoidance of such interference, please see ELECTROMAGNETIC
COMPATIBILITY INFORMATION. It is suggested that the blood pressure
monitor be kept 10 meters away from other wireless devices, such as WLAN
unit, microwave oven, etc.
9. This product should not be used as a USB device.
10.This product is verified by auscultatory method. It is recommended that you
check Annex B of ANSI/AAMI SP-10:2002+A1:2003+A2:2006 for verification
method details if needed.
11.If the determined blood pressure (systolic or diastolic) is outside the rated
7