CALIBRATION
INSTRUCTIONS
The sensitivity of the leakage test circuit in the Model
667 is adjusted
at the
factorv
so
that
a leakage
of
100
megohms
just
reads
in
the
GRID
EMISSION—REJECT
area of the meter
scale.
Should this sensitivity
control
require
readjustment,
proceed
in the fol-
lowing manner:
1.) Remove main panel from case, after removing screws from panel and
center of bottom
of carrying case.
(Between
molded feet)
2.) Place a 100 megohm
resistor in tube socket #6 between pin
1 and 9.
3.) Rotate SELECTOR
switch
to position #1.
4.)
Depress
the
LEAKAGE
pushbutton
and
observe
the
meter.
Adjust
the white trimpot on the P.C. board adjacent to the meter until the
meter
pointer
reads
at the division
of the
GREEN
and
the
GRID
I MISSION—REJECT
sections
of the scale.
The shorts test circuit is factory adjusted to a sensitivity of one megohm.,
Should this circuit require readjusting, proceed in the following manner.
1.)
Remove
main
panel
from
case as above.
2.) Place a one megohm
resistor in tube socket #6 between pin 1 ] and 9.
3.) Rotate the SELECTOR
switch to position #1.
4.) While
observing
the SHORTS
light,
adjust
the
blue
trimmer
po-
tentiometer on the P.C.
board
until the SHORTS
light just barely
glows.
5.)
Reassemble
tester.
12