Verify operation
Operation can be verified using four methods at the device via the LCD display,
using the Field Communicator, at the Gateway via the Gateway's integrated
web server, or using AMS
3.1
LCD display
During normal operation, the LCD display will show the PV value at the
confirmed update rate. Refer to the
LCD display messages. Select the Diagnostic button to display the TAG, Device
ID, Network ID, Network Join Status, and Device Status screens.
Searching for
network
n e t w k
A - s r c h
3.2
Configuration
For HART Wireless transmitter communication via a Field Communicator or
AMS Device Manager, a Rosemount WDM Device Dashboard (DD) is required.
Contact the Rosemount Wireless Specialist Team
([email protected]) to obtain the DD.
To provide accurate sensor readings, values from each sensor data sheet need
to be configured and stored in the Rosemount WDM Transmitter. Navigate to
Configure > Manual Setup > Sensor > Sensor Setup > Data Sheet Values.
The special configuration items the user will need to enter are those provided
(i.e. Range, Supply Voltage, 0%, 50%, and 100%). These values can be found in
the custom data sheet sent with each sensor.
6
™
Wireless Suite or AMS Device Manager.
Reference Manual
Joining network
n e t w k
J O I N G
for error codes and other
Connected with
one parent
n e t w k
1 P A R N T
February 2019
Connected with
two parents
n e t w k
2 P A R N T