GaleMed 2163 Kullanım Kılavuzu - Sayfa 4

Solunum Ürünleri GaleMed 2163 için çevrimiçi göz atın veya pdf Kullanım Kılavuzu indirin. GaleMed 2163 11 sayfaları.

Principles Of Operation 3 
The Infant and Child models of the MR‐100™ Plus    Manual Resuscitator are 
fitted with pressure relief valves (see figure 4). These valves open when the 
pressure is in excess of 40±5cmH
 
 
 
 
 
Figure 4. Pressure Relieve Valve. 
 
Should higher inspiratory pressure be required, the pressure relief valve 
may be overridden by placing thumb over the valve as shown in figure 5. It 
Is also possible to lock the pressure relief valve in this position with Lock 
type Non rebreathing Valve. Just twist the valve half a turn and it will stay 
in the overridden position. (Lock Arrow forward to the NRV direction as 
Figure 5.1). 
 
 
 
 
 
 
Figure5.1 
Figure 5. Overriding the Pressure Relieve Valve 
 
 
 
0. 
2
Operating Instruction 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Note: lf the patient is intubated or has a tracheostomy, the resuscitator can be 
connected by removing the mask and attaching the non-rebreathing valve outlet   
directly to the connector of the endotracheal or tracheostomy tube. 
1.
Position victim face up. 
 
2.
Clear foreign objects 
out of patient's mouth 
and throat. 
 
3.
Insert oropharyngeal 
airway (if available), in 
accordance with the 
manufacturer's 
directions, to open the 
patient's mouth and 
prevent the tongue 
from obstructing the 
airway. 
 
4.
Position yourself behind 
the patient's head, 
 
extend the head back 
and pull the chin