9. Hold the handpiece in a VERTICAL upright position and step on the
foot switch. Within 10-15 seconds water should come out of
the open end of the handpiece IF NOT, QUICKLY
turn WATER KNOB clockwise to get water flow,
some air will escape at the same time. This step is
IMPORTANT because trapped air will cause HOT
SPOTS (usually noted by a loud screeching sound)
which can affect the performance of the
AUTOSCALER® or permanently damage the
handpiece assembly.
10. Inserts and handpiece will burn in 45-60 seconds in the absence of
water. A (18cc—30cc/+/min) flow is the normal flow range, and
necessary at all times. NEVER touch the Cavitating/Vibrating end of the
insert tip to any flesh such as the cheek, tongue, or gums unless
performing special periodontal procedures. This will cause a FRICTION
BURN to the patient. Always keep the insert in motion, do not rest the
handle or insert on the lips during a procedure. The insert tip should
touch primarily the calculus on the tooth surface and NOT the enamel
of the tooth surface.
11. Allow the air bubbles to escape for about 10-20 seconds. Turn the
POWER KNOB to the RIGHT in the Medium position and the insert tip
should Cavitate/Vibrate with a strong mist. Turn the water knob to
the right if there is no mist or weak misting.
12. The application of the power knob is for controlling the Output Power
to the insert tip in a smooth linear action. The operator can select the
EXACT power needed for cleaning. Set power knob LOW (Left Direction)
to MEDIUM (Center) for cooler handpiece/insert temperature.
13. The high power setting is for applications where an excessive amount
of calculus is built up. The Low power setting can be used for
periodontal Subgingival curettage and root planing procedures. Most
scaling should be done only on the LOW to MEDIUM power setting.
14. If SLIMLINE™, FSI™ or Modified Periodontal inserts are used set the
POWER Knob in the BLUE shaded area. This power level is designed to
be used with these fragile specialty inserts. Inserts may give a drip and
mist combination, this is normal for these inserts. This is actually
preferred by most Periodontal professionals due to the water drip
being ultrasonically activated within 0.5mm of the tip assisting in killing
plaque.
UNIT IS NOW READY FOR USE
9

SHUT DOWN PROCEDURE

1. IMPORTANT: At the end of the work day turn off the water supply to
the AUTOSCALER® by disconnecting the QUICK RELEASE VALVE water
connector or shutting off the main operatory water supply. This will
prevent water damage to your office by aging or unseen defects in
water lines that could rupture under continuous pressure.
2. Purge water from the AUTOSCALER® water hose and plumbing system
at the end of the day.
Connect the QUICK RELEASE VALVE water connector to an air
source.
Turn the Power Knob to the Left to the lowest setting and the
Water Knob to the Right to the highest setting.
Press the Toggle switch to the ON-I position
Hold the handpiece assembly over a basin and step on the
footswitch and run the scaler just until all water is expelled via
the handpiece assembly.
3. Press the Toggle Switch to the OFF-O position. The light should be off.
4. Reconnect the QUICK RELEASE VALVE water connector to the water
supply.
5. REMEMBER to reconnect (turn on) the water the next day before
usage.

TROUBLESHOOTING

If the unit does not function, proceed with the following:
1. If the power L.E.D. light above the switch is not lit, replace the FUSE
on the rear panel with a "LITTLE FUSE" 2 amp Fast Blow style. If the
fuse opens again DO NOT attempt to replace the fuse as this indicates
a serious condition that must be examined by the factory. Call for
service (800) 648-9445.
2. Check the plug. Is it all the way in the receptacle?
3. Try another insert. Make sure it is the correct frequency, 25K or 30K,
depending on your model. Each model has a frequency (25K or 30K)
label on the cover and the frequency is also indicated on the serial
number plate located on the back of the device. The most common
mistake when purchasing a new unit is selecting the wrong
frequency insert.
4. If water flow is insufficient, check water pressure at the operatory for
minimum of 20 P.S.I. The Insert Water slot could be clogged with
debris. Try another insert. See if the QUICK RELEASE VALVE water
connector is locked fully in place.
5. If the handpiece handle or insert operates HOT: REDUCE POWER
LEVEL and INCREASE WATER FLOW.
10