2 Technical data

Feature
Absolute pressure sensor
(absolute)
Vacuum measuring range
Sensor overload (relative)
Vacuum resolution
Vacuum accuracy
(at 22 °C, after field calibration,
confidence level 95%)
Operating temperature
Storage temperature
Temperature measuring range -20 to 50 °C / -4 to 122 °F
Temperature resolution
Battery life
Protection class
Parameter
Measuring cycle
Sensor
Connections
Warranty
Values
max. 6 bar (87 psi)
Pressure exceeded.
Destruction of the absolute pressure
sensor!
- Do not exceed the maximum value.
1100 - 0 mbar / 825080 - 0 micron
5 bar / 72 psi
0.01 hPa / 10 micron
- 0 to 1.33 hPa / 0 to 1000 micron: up to
±10 micron
- 0 to 200 hPa / 0 to 150000 micron:
±0.3% FS = ±0.6 hPa
- 200 to 1100 hPa / 150000 to 825080 micron:
±0.3% FS = ±3.3 hPa
-20 to 50 °C / -4 to 122 °F
-20 to 50 °C / -4 to 122 °F
0.1 °C / 0.1 °F
2400 h (2x AA) (approx. 130 h with background
illumination activated)
IP 42
mmHG, Torr, mbar, hPa, micron, inH2O, inHg.
Pa
0.5 sec
1× absolute pressure sensor
- 2× 7/16" UNF
- 1x MiniDIN
2 years
2 Technical data
ATTENTION