19.

GUIDANCE AND MANUFACTURERS DECLARATION

FCC:
FCC ID: 2AHYU-9990005001
IC: 21382-9990005001
FCC Part 15.19(a):
This device complies with part 15 of the FCC Rules. Operation is subject
to the following two conditions: (1) This device may not cause har mful
interference, and (2) This device must accept any interference received,
including interference that may cause undesired operation.
FCC Part 15.21:
Changes or modifications not expressly approved by the party
responsible for compliance could void the user's authority to operate the
equipment.
FCC Part 15.105(b):
This equipment has been tested and found to comply with the limits for
a Class B digital device, pursuant to Part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against har mful
interference in a residential installation. This equipment generates uses
and can radiate radio frequency energy and, if not installed and used
in accordance with the instructions, may cause har mful interference to
radio communications. Howev er, there is no guarantee that interference
will not occur in a particular installation. If this equipment does cause
har mful interference to radio or television reception, which can be
deter mined by tur ning the equipment off and on, the user is encouraged
to try to correct the interference by one of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that
to which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
ISED RSS-Gen Notice:
(1) This device may not cause interference; and (2) This device must
accept any interference, including interference that may cause undesired
operation of the device.
1) l'appareil ne doit pas produire de brouillage; 2) l'appareil doit
accepter tout brouillage radioélectrique subi, même si le brouillage est
susceptible d'en compromettre le fonctionnement.
For support, please contact at +1-650-397-6103 or visit www.dynosense.com
All manuals and user guides at all-guides.com
18

Electromagnetic Emissions

The ME Equipment is intended for use in the electromagnetic environment
specified below. The customer or the user of the ME Equipment should
assure that it is used in such an environment.
Emission Test
Compliance Electromagnetic environment -
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Har monic emissions
Class A
IEC 61000-3-2
Voltage
Complies
fluctuations/flicker
emissions IEC
61000-3-3
Radiated emissions limits and results:
EN/IEC 60601-1-2 Class B Limits below 1GHz
Frequency
10m measuring distance 3m measuring distance
range MHz
Quasi-peak
dB(μV/m)
30 to 230
30
230 to 1000
37
On a test site, class B equipment can be measured at a nominal
distance of 3m or 10m. A measuring distance less than 10m is allowed
only for equipment which complies with the definition given in 3.10.
At the transition frequency, the more stringent limit shall apply.
a
The limits specified for the 3m separation distance apply only to
small equipment meeting the size criterion defined in 3.10.
For support, please contact at +1-650-397-6103 or visit www.dynosense.com
guidance
The device uses RF energy only for
its inter nal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
The device is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
a
Quasi-peak
dB(μV/m)
40
47
19