Cabinet Cleaning -The surfaces of the enclosure can be cleaned with a solution of mild soapy water or Windex™ type glass
cleaner.
Under normal use the blower unit, which has a sealed bearing, should provide many years of uninterrupted use. There are no user
serviceable parts in the blower unit housing. The filters should be checked for performance and replaced on a regular basis.
There is a risk that internal parts of model ES-7000 blower unit may be contaminated with whatever substance the filter is intend-
ed to remove. When replacing filters please follow appropriate guidelines for handling materials the enclosure is being used to con-
tain. Wear OSHA approved gloves, safety glasses and filtered breathing mask.
Used filters should be placed in impermeable bags for disposal in accordance with local regulations.
For further information, call or write the Occupational Safety and Health Administration. Department of Labor, Washington, DC
(410) 865-2055 or your local health department.
Turn off blower unit and remove line cord from blower. Unsnap latches on either side of blower unit and lift blower unit from filter
base. Loosen wingnuts to allow filter removal.
Remove old filters and place in heavy plastic bag. Follow reverse order of these instructions when installing new filters. The gas-
keted side of the filter should always be facing downward. The filter replacement history should be maintained using the card sup-
plied in the clear pouch on front of the blower.
Airfiltronix warrants this product for one year from data of purchase. We will repair without charge, or replace at our discretion, any
defects due to faulty material or workmanship. Please return the unit in original packaging or suitable equivalent, transportation
prepaid and insured directly to us. This warranty does not apply to filters, accessories or damage caused by improper repair or
unauthorized repairs attempted by anyone other than by manufacturer.
Regular Maintenance
Safety Procedures
Filter Maintenance and Replacement
Warranty
For the name of your local representative contact:
AIRFILTRONIX
CLEAN AIR FOR INDUSTRY MEDICINE & RESEARCH
154 Huron Avenue, Clifton, New Jersey 07013
800-452-8510 • PHONE: 973-779-5577 • FAX: 973-779-5954
www.airfiltronix.com • e-mail: [email protected]