Decagon AquaLab Hızlı Başlangıç Kılavuzu - Sayfa 6

Ölçüm Cihazları Decagon AquaLab için çevrimiçi göz atın veya pdf Hızlı Başlangıç Kılavuzu indirin. Decagon AquaLab 6 sayfaları. Vapor sorption analyzer

Decagon AquaLab Hızlı Başlangıç Kılavuzu
V S A Q u i c k S t a r t G u i d e
2. AquaLab VSA Performance Verification overview—
  •  AquaLab VSA performance and cleanliness should be
    checked frequently; before each isotherm test is 
    recommended.
  •  Press   until you are at the Measurement screen (2nd tab).
  •  Use AquaLab Verification Standards to verify performance
    or adjust linear offset.
  •  To verify, remove weigh pan and then place sample cup
    containing one of AquaLab's verification standards in
    sample chamber. Close lid and carefully avoid spillage; See
    Figure 5: Sample Chamber.
  •  Move latch left and/or press   to start a reading. The
    pinwheel will spin indicating a measurement is running.
a
    The 
 reading for the standard and the temperature
w
    will display when finished. Make sure the 
    on the standard is within +/-0.005 of the correct 
    standard. (+/- 0.02 if cap sensor is used)
  •  If an offset adjustment is needed, refer to chapter 7 in the
    user's manual.      
Base
Figure 5: Sample chamber
a
 reading
w
a
 for the
w
Lid
3. Sample Preparation—
  •  Fill sample cup no more than half full (~7ml). 
Do Not Overfill!
  •  Minimum sample amount should cover bottom of sample
    cup (500-2000 mg) (small gaps OK).
  •  Ensure rim and outside of sample cup are clean.
  •  Make sure sample temperature is not more than 4°C above
    chamber temperature.
4. Starting a Test—
  •  Before starting a test, set the Volatiles: Yes/No option in
    the configuration menu. If your sample has volatiles,
    select "Yes" otherwise select "No".
  •  On the main screen, select the test wizard  .
  •  Provide a name for the test using the arrow keys and then
    press  . (The name of a previous test will appear by
    default.)
  •  Enter initial moisture content if known (optional). Press 
    to continue or   to go back.
  •  Setup each stage of the isotherm test. A stage is used to
    adjust the isotherm settings for an individual sample (up to
    20 stages). Each stage is setup with the isotherm method 
    (DDI or DVS) and the settings for that portion of the
    isotherm. Once a stage is created, it can be modified or
    deleted anytime before the test is started. 
i.e. For milk powder, Stage 1 could use the DDI method to
ramp quickly from 0.1 to 0.4
the DVS method to step from 0.4 to 0.8
a
0.05
increments.
w
  •  To setup a new stage, select "new" and press  . A screen
    with several parameters will appear. If a previous test
    has been run, the stages for that test will be listed by
    default. These stages can be used as is or updated with
    new settings. If you need to delete a stage, go to the
    specific stage and press  . Scroll down to the bottom of
V S A Q u i c k S t a r t G u i d e
a
and Stage 2 could use
w
a
slowly in
w