Major tech K2007R Manuel
Ölçüm Cihazları Major tech K2007R için çevrimiçi göz atın veya pdf Manuel indirin. Major tech K2007R 2 sayfaları. 1000a ac clamp meter
K2007R
1000A AC Clamp Meter
Current Sensor
Barrier
Function Switch
Hold Key
LCD
COM Terminal
V/Ω Terminal
Wrist Strap
Barrier of Test Leads
Cap
Safety Warnings
This instrument has been designed, manufactured and tested
according to IEC 61010: Safety requirements for Electronic
measuring apparatus, and delivered in the best condition after
passed the inspection. This instruction manual contains warnings and
safety rules which must be observed by the user to ensure safe
operation of the instrument and retain it in safe condition. Therefore,
read through these operating instructions before using the
instrument.
WARNING
●
Read through and understand the instructions contained in this
manual before using the instrument.
●
Keep the manual at hand to enable quick reference whenever
necessary.
●
The instrument is to be used only in its intended applications.
●
Understand and follow all the safety instructions contained in
the manual.
●
It is essential that the above instructions are adhered to.
Failure to follow the above instructions may impair the
protection provided by the instrument and test leads, and may
cause injury, instrument damage and/or damage to equipment
under test.
The symbol
indicated on the instrument means that the user
must refer to the related parts in the manual for safe operation of the
instrument. It is essential to read the instructions wherever the
symbol
appears in the manual.
DANGER is reserved for conditions and actions
that are likely to cause serious or fatal injury.
WARNING is reserved for conditions and actions
Trigger
that can cause serious or fatal injury.
CAUTION is reserved for conditions and actions
that can cause injury or instrument damage.
● Marks listed below are used on this instrument.
User must refer to the manual.
Instrument with double or reinforced insulation
Indicates that this instrument can clamp on bare
conductors when measuring a voltage corresponding
to the applicable measurement category, which is
marked next to this symbol.
AC
DC
Ground (Earth)
This instrument is subject to WEEE Directive
(2002/96/EC).
Please contact our dealer near you at disposal.
Measurement Category
O Circuits which are not directly connected to the mains
power supply.
CAT II Primary electrical circuits of equipment connected to an AC
electrical outlet by a power cord.
CAT III Primary electrical circuits of the equipment connected
directly to the distribution panel, and feeders from the distribution
panel to outlets.
CAT IV The circuit from the service drop to the service entrance,
and to the power meter and primary over current protection
device(distribution panel).
This instrument is designed for CAT IV 300V / CAT III 600V. Test
leads K-7066A with the supplied cap are designed for CAT IV 600V /
CAT III 1000V and without the caps are for CAT II 1000V.
Interior wiring
●
Never make measurements under the circumstances exceeding
the designed measurement category and the rated voltage of
the instrument and the test leads.
●
Do not attempt to make measurement in the presence of
flammable gasses. Otherwise, the use of the instrument may
cause sparking, which can lead to an explosion.
●
Never attempt to use the instrument if its surface or your hand
is wet.
●
Do not exceed the maximum allowable input of any measuring
range.
●
Never open the Battery cover during a measurement.
●
To avoid electrical shock by touching the equipment under test
or its surroundings, be sure to wear insulated protective gear.
●
Never measure current while the test leads are inserted into the
input terminals.
●
Test leads to be used for voltage. Measurements shall be rated
as appropriate for Measurement Category III or IV according to
IEC 61010-031 and shall have a voltage rating of 600V or
higher.
●
Barriers on the instrument body and the test leads provide
protection to keep your fingers and hands from touching an
object under test. Keep your fingers and hands behind the
barriers during measurement.
WARNING
●
Never attempt to make measurement if any abnormal
conditions, such as broken case and exposed metal parts are
found on the instrument or test leads.
●
Verify proper operation on a known source before use
or take action as a result of the indication of the instrument.
●
Firmly attach the caps to the test leads when performing
measurements in CAT III or higher test environments. When
K2007R and the test leads are combined and used together,
whichever is lower category & voltage to earth either of them
belong to is applied.
●
Do not rotate the Function Switch if the instrument and the
equipment under test are connected.
●
Do not install substitute parts or make any modification to the
instrument. For repair or re-calibration, return the instrument to
your local Major Tech distributor
CAUTION
●
Use of this instrument is limited to domestic, commercial and
light industry applications. Strong electromagnetic interference
or strong magnetic fields, generated by large currents, may
cause malfunction of the instrument.
●
Connect the test leads to the terminals firmly.
●
This instrument isn't water proofed. Keep away from water.
●
Do not pull or twist the test leads to prevent the risk of
damage.
●
Power off the instrument after use. Remove batteries if the
instrument is to be stored and will not be in use for a long
period.
●
Do not expose the instrument to the direct sunlight, high
temperature and humidity or dewfall.
●
Use a cloth dipped in water or neutral detergent for cleaning the
instrument. Do not use abrasives or solvents.
NOTE
●
The LCD shows some digits at the AC V and the DC V
O: Device which is not
directly connected to
ranges even while the test leads are open. In addition,
the mains power supply
the LCD shows some digits instead of 0 when short-
circuiting the test leads. However, these phenomena
don't affect measurement results.
1. Specifications
DANGER
Temperature: 23 ± 5°C, Humidity: 45 - 75%
ACA
Range
Display Range
600A
0.0, 0.4-629.9A
1000A
570-1049A
*1 : when the measured conductor is in the center of clamp
jaw Guaranteed accuracy : 0.8A -1000A,less than
1500A Peak Input protective current :AC1200A
ACV
Range
600V
0.0, 0.6-629.9V
Guaranteed accuracy : 1.0V - 600.0V,less than 900V Peak Input
protective voltage :AC/DC720V 10 sec
DCV
Range
600V
Guaranteed accuracy : 0.0V - ± 600.0V
Input protective voltage :AC/DC720V 10 sec
ACV/DCV input impedance : approx. 10MΩ
Resistance / Continuity
Range
600Ω
0.0-629.9Ω
6kΩ
0.570-6.299 kΩ
Cont.
0.0-629.9Ω
Guaranteed accuracy : 0.0Ω - 6.000kΩ
Open-loop voltage : 3V or less
Input protective voltage :AC/DC600V 10 sec
Measuring method : ⊿Σ modulation ●Over-range indication : OL
●
●
Measurement cycle : 2.5 times per second
●
Crest factor : less than 3 (45-65Hz)
Add ±0.5%rdg±5dgt to above specified accuracies. (Applicable
functions :ACA,ACV)
●
Applicable Standards :
IEC 61010-1/ 61010-2-032/ 61010-2-033 (instrument) Pollution
degree 2, Indoor use,Altitude up to 2000m CAT III 600V / CAT IV
300V
IEC 61010-031(Test leads Model 7066A)
w/ caps CAT IV 600V / CAT III 1000V w/o caps CAT II 1000V
IEC61326 (EMC), EN50581 (RoHS)
●
Withstand voltage: AC5160Vrms 5sec
between current sensor and enclosure or circuit and enclosure
●
IP rating : IP40 (IEC60529) ●Insulation resistance :
>100MΩ /1000V between enclosure and electrical circuit
●
Operating Temperature and humidity range :
0 to 40°C 85%RH or less (no condensation) ACV
●
Storage Temperature and humidity range :
-20 to 60°C 85%RH or less (no condensation)
●
Power source : DC3V R03 / LR03 (AAA) ×2
●
Current consumption : < 4mA
Accuracy (sine wave)*1
±1.5 %rdg±4dgt (45-65Hz)
±2.0 %rdg±4dgt (40-400Hz)
Display Range
Accuracy (sine wave)
±1.2 %rdg±3dgt (45-65Hz)
±1.5 %rdg±4dgt (40-400Hz)
Display Range
Accuracy
0.0-±629.9V
±1.2%rdg±3dgt
(Auto-range)
Display Range
Accuracy
±1.3%rdg±5dgt
±2.0%rdg±3dgt
Bz threshold value <90Ω