4.8. Direction of the sensor head:
There is a mark on the top of the "Sensor Head", when making the
measurement, this mark should be against the measured wind - refer to
Fig. 4 and Fig. 5. When the sensor head face is against the measurement
air, then sensor head face against the measured air, then the upper
display will show the air velocity value. The lower display will show the
temperature value.
5. PERFORMING A MULTI-POINT MEAN CALCULATION
• Press "Mean".
Mean is lit. The number of readings recorded is displayed in the upper
line, while the current reading is displayed in the lower line.
• To change between displaying the temperature, flow velocity and
calculated. Volumetric flow rate: Press "Flow/Temp".
• If we want to change the units of the current reading, press "Unit
• To include readings (in the desired quantity): Press "Enter" (several
times).
• To end measurement and calculate the mean value: Press "Mean".
Mean flashes. The calculated spot mean value is displayed.
• To return to measurement view: Press "Mean".
".
9