Apply a reflective mark to the object being measured. Hold down the MEAS
button and align the visible light beam with the applied target. Verify that
monitor indicator lights when the target aligns with the beam. The meter turns
on with the previously used mode. If you need to change the mode, release the
MEAS button and press the MODE button to change between RPM and REV.
Hold down MEAS to activate the new mode. The instrument auto powers off in
10 sec after releasing the MEAS button. Press the MEM button to recall the
last measurement readings for the Max, Min and Last value.
Measuring Considerations
1.
Reflective Mark
• Cut the adhesive tap provide into approx. 12mm (0.5") squares and
apply one square to each rotation shaft.
• The non-reflective area must always be greater than the reflective area.
• If the shaft is normally reflective, it must be covered with black tape or
black paint before attaching reflective tape
• Shaft surface must be clean and smooth before applying reflective tape.
2.
Very Low RPM Measurement
• As it is easy to get high resolution and fast sampling time, if measuring
the very low RPM values, suggest user to attach more "REFLECTIVE
MARKS" averagely. Then divide the reading shown by the number of
"REFLECTIVE MARKS" averagely. Then divide the reading shown by the
number of "REFLECTIVE MARKS" to get the real RPM
160 x 58 x 39mm