Major tech MT800 Kullanım Kılavuzu - Sayfa 7

Multimetre Major tech MT800 için çevrimiçi göz atın veya pdf Kullanım Kılavuzu indirin. Major tech MT800 8 sayfaları. Analogue multimeter

12.REPLACEMENT OF FUSE
WARNING: To avoid electric shock, disconnect measuring terminals
before removing fuse.
CAUTION: for continued protection against fire, replace only with 0.3A,
250V fuse.
1. Use a phillips screwdriver to remove the screw on the back cabinet. Pull
apart the front and back cabinets.
2. Replace the fuse with a new fuse of the same type / rating (0.3A, 250V)
3. Do not touch any inside area other than the fuse.
13. TIPS FOR BEST HANDLING
1. Always observe correct polarity: red lead to positive and black lead to
negative terminal.
2. When checking unknown voltages, use the highest range first and
gradually decrease until readings are obtained.
3. For exact readings, look at the scale from the point where the pointer and
its reflection in the mirror come together, otherwise a reading error due
to parallax will result.
VR1
R12
R1 R2 R3 R4
R5
25 10 50 250 500 X10 XIK 500 250 50 10
DCV
+
Ø
R1. 4.4K
R2. 15K
R3. 80K
R4. 400K
VR2
185uA
D
D
3
2
BATT
Ω
ACV
R5. 500K
R9. 46Ω
R6. 1.14K
R11. 9Ω
R7. 2 Ω
R12. 3K
R8. 2Ω
R13. 36Ω
R13
R8
R7
R9
R10
R11
R6
9 1.5 250
10
500
m
m
μ
-
BATT
DCA
Ø
D1,D3.
IN4148
D2. IN60P
F0.5A
7
7