Avante Volumax VOS Hızlı Referans Kılavuzu - Sayfa 3
Evcil Hayvan Bakım Ürünü Avante Volumax VOS için çevrimiçi göz atın veya pdf Hızlı Referans Kılavuzu indirin. Avante Volumax VOS 4 sayfaları.
Avante Volumax VOS
Quick Reference Guide
ALARM ICON
GENERAL FAULT
(FLASHING)
GENERAL FAULT
(FLASHING)
DISPLAYS DASHES ( - - )
BLOCKED TUBE
(FLASHING)
WATER OUT
(FLASHING)
DISPOSABLE WATER
PATH (FLASHING)
BATTERY CHARGING
(FLASHING)
BATTERY (FLASHING)
2601 Stanley Gault Pkwy, Suite 101
800-923-8413
502-244-6345
•
INDICATES
Malfunction of sensor or
control system
% O₂
High back pressure
No water in disposable
continues without heat-
ing or water circulation.
Disposable water path
faulty or not detected.
Unit will not run.
The internal battery backup is not fully charged. The unit would not run on battery for
the full rated time in the event of a power failure. No action is necessary.
The unit is running in
and blending continues
without heat or water
circulation.
GENERAL FAULT ALARMS:
percentage will cause a General Fault alarm indicated by this icon. Unit will continue gas delivery in a fault condition
until placed in Standby. With the exception of O₂ sensor replacement, the unit must be repaired by an approved
ser
AC power and then press the Run/Standby Button.
Louisville, KY 40223 USA
•
FAX: 502-244-0369
•
CAUSE
Internal component
Cannot be corrected by user:
failure
send for service.
Depleted or defective
Obstructed or kinked
Clear obstruction, check cannula
cannula/delivery tube,
type, re-install DPC
incorrect cannula for
improperly seated
Sterile water empty,
Disconnect patient. Replace water
or obstructed inlet
bag or straighten inlet tube.
tube.
Restart unit.
Disposable water path
If disposable water path is
defective, not properly
present, remove and replace to
seated or not installed.
reset detector.
AC power is
Reconnect AC power.
disconnected
www.avantevet.com
ACTION