Instructions for Use for KardiaMobile 6L
(2) This device must accept any interference, including interference that may cause undesired
operation of the device.
Ingress Protection Marking
KardiaMobile 6L is IP22 rated. KardiaMobile 6L is protected against insertion of fingers and is
not affected by vertically dripping water. KardiaMobile 6L has been tested with relevant
requirement standard IEC 60601-1-11:2015.
Applied Parts
The 3 electrodes (Left Hand Electrode, Right Hand Electrode, and Left Leg Electrode) are Type
CF Applied Parts.
Operational temperature of the device is +10°C to +45°C. If ambient temperature exceeds
+41°C, Applied Parts can exceed +41°C.
Troubleshooting
If you experience difficulties using your KardiaMobile 6L, refer to the troubleshooting guide
below or contact technical support at [email protected].
I'm having trouble getting a clear recording.
o Clean the electrodes using a damp soft cloth. Wash your hands with soap and water. Use
a small amount of water to moisten the skin where your fingers make contact with the
electrodes.
o If recording a Six-Lead EKG, it is important to place device on your left leg (knee or inside of
the ankle). The device must be used on bare skin for an accurate recording.
o Ensure that your arms, hands and left leg remain still to reduce muscle noise. Do not apply
too much pressure to the electrodes.
○ Avoid close proximity to items that may cause electrical interference (electronic equipment,
computers, chargers, routers, etc.)
○ If you wear hearing aids, turn them off prior to recording.