A-M Systems 3000 Kullanım Kılavuzu - Sayfa 10
Amplifikatör A-M Systems 3000 için çevrimiçi göz atın veya pdf Kullanım Kılavuzu indirin. A-M Systems 3000 18 sayfaları. Ac/dc differential amplifier
MODE (STIM-GATE-REC)
Ω
TEST
3. Turn on power to the Model 3000 and allow it to warm up for 5 minutes.
4. Non Headstage option: connect measuring electrode to the black wire and a
reference electrode to the white wire of the probe cable.
Headstage option: connect measuring electrode to the +
input of the head stage and a reference electrode to the -
input of the head stage (see diagram to the right). Note: the
reference electrode can be tied to the gnd input on the head
stage. Clamp the Head stage in a micromanipulator.
5. Dip the measuring and reference electrode into a beaker of
physiological saline solution (or the solution in which the
tissue will be bathed). The solution should have the same
temperature and ionic strength as that in which
measurements will be made. Note: immerse the measuring electrode to
approximately the same depth as will be used during the measurement.
6. Connect an oscilloscope to the
2ms/division. Zero the scope. Flip the input switch to
7. Observe the offset potential between the two electrodes on the oscilloscope. Set
the
DC OFFSET (+ OFF -)
knob to zero the amplifier output. Note: make sure your oscilloscope is in DC
mode.
8. Turn on the Ω
electrode.
9. Adjust the oscilloscope for a good display of the square-wave. The Oscilloscope
will display a square wave at 0.1 V peak to peak for every 1 MΩ of electrode
resistance.
10. Increase the
CAPACITY COMPENSATION
Avoid overcompensation, which will cause ringing, excessive noise, and high
frequency oscillation.
11. Turn the
CAPACITY COMPENSATION
present position and the off position. This will minimize the chance of
overcompensation during recording.
12. Turn off the Ω
13. Apply the electrodes to the experimental preparations.
14. Apply the
HIGH PASS, LOW PASS,
15. Increase the gain switch until the output voltage is in the range necessary for
saving data. Note: The
A-M Systems
OUTPUT
switch to the appropriate polarity and adjust the
button to inject a 100 Hz square-wave current through the
TEST
switch to stop the test signal.
TEST
and
Switch should be set so that the signal at the
GAIN
131 Business Park Loop, P.O. Box 850 Carlsborg, WA 98324
Telephone: 800-426-1306 * 360-683-8300 * FAX: 360-683-3525
E-mail: [email protected] * Website:
REC
OFF
BNC, with the horizontal sweep rate set to
DIFF.
to "square-up" the corners of the waveform.
knob to approximately half way between its
if necessary.
NOTCH FILTERS
http://www.a-msystems.com
DC OFFSET
OUTPUT
7