acaia Cinco Kullanım Kılavuzu - Sayfa 13
Terazi acaia Cinco için çevrimiçi göz atın veya pdf Kullanım Kılavuzu indirin. acaia Cinco 18 sayfaları.
Note
1. In Auto Start Mode, the timer will stop when the espresso flow stops
2. The auto-tare function is performed when the weight exceeds 2 grams.
3. If you do not operate the scale for 14 seconds, it will reset the timer and perform tare operation.
4. In Auto Start Modes, the timer will also stop when a negative weight (i.e. removal of cup) is
detected.
*To Enable/Disable Auto Start Mode, please go to Chapter 5: Configuration.
Chapter 4: Calibration
Calibration Mode
1.
While the scale is in Weighing Mode and the weighing unit is in gram, press
until LED display shows [ CAL ] and then [ 100 ].
2. Place a
100g
calibration weight on scale and wait for the scale to execute calibration procedure.
3. The LED display shows [
the calibration weight when completed.
*
Note: Calibration weight of 50g,100g, 200g, or 500g can also be used.
Chapter 3: Basics
_End_
] when the calibration procedure is completed. You may unload
T
Button
rapidly
13