Allied Healthcare EPV200 Kullanıcı Kılavuzu - Sayfa 4

Tıbbi Ekipman Allied Healthcare EPV200 için çevrimiçi göz atın veya pdf Kullanıcı Kılavuzu indirin. Allied Healthcare EPV200 20 sayfaları. Portable battery powered ventilator with oxygen source required

Allied Healthcare EPV200 Kullanıcı Kılavuzu
5

Specifications

A.
Gas Supply Pressure: 280 kPa (40.6 psi) to 600 kPa (87 psi) oxygen DISS
B.
Breaths Per Minute (BPM): Accuracy: ±10%
BPM Range: 1 second inspiratory time = 0 and 5 to 30
BPM Range: 2 second inspiratory time = 0 and 5 to 20
C.
Tidal Volume (Tv): Accuracy: ±10% with 100% oxygen
Tidal volume range 1 second inspiratory time = 200 ml to 600 ml
Tidal volume range 2 second inspiratory time = 400 ml to 1200 ml
D.
Inspiratory Time (I
1 second or 2 second selection
E.
Safety Pressure Relief: Fixed at a maximum of 60 cm H
F.
Low Source Gas Alarm: Activates at 275 to 241 kPa (40 to 35 psi) source
pressure.
G.
High Airway Pressure Alarm Electronic: Activates at 45 cm H
Alarm sound level is greater than 58 decibels.
H.
Breath Delivery Alarm: Activates if the airway pressure does not go
above 9 cm H
detected.
I.
Battery Life: Greater than 48 hours at room temperature (10 BPM and
2 seconds I
)
t
J.
Oxygen Inlet Filter: 65 micron sintered bronze
K.
Leakage: The unit shall be designed so that pressurized oxygen is not allowed
to leak through any seals or fittings.
L.
Gauge: 0-99 cm H
M.
Inspiratory and Expiratory Resistance: 5 cm H
N.
Inadvertent PEEP: < 2 cm H
O.
Inadvertent Continuing Expiratory Pressure: < 2 cm H
P.
Dead Space: < 5.5% of minimum tidal volume
Q.
Weight: 1.4 kg (3.1 lbs.) with batteries
R.
Size: 3.5 x 7.0 x 9.3 (88.9 x 177.8 x 236.2 mm)
S.
Operating Conditions: -9 to 50°C (15 to 122°F) 5% - 95% RH non condensing
T.
Storage Conditions: -40 to 60°C (-40 to 140°F) 5% - 95% RH non condensing
U.
Shipping Conditions: -40 to 60°C (-40 to 140°F) 5% - 95% RH non condensing
Breathing mode: Assist Control with ≤2 cm H
V.
spontaneous breath.
Latex Free: This product does not contain Latex.
): Accuracy: ±10%
t
O during a span of 15 seconds, or if a spontaneous breath is
2
O accuracy ± 5% or 1 cm H
2
O
2
O, whichever is greater
2
O (.5kPa) maximum
2
O required to trigger a
2
3
O
2
O.
2
O
2