4.5 TO REMOVE CASE TOP FROM CASE BOTTOM:
1.
Always disconnect any power source.
2.
Remove the screws from the sides and back of the unit.
3.
Remove the Case Back from the Case Front and lay along side the Case Bottom.
4.
Disconnect the ground wire from the bulkhead assembly.
5.
Disconnect the tubing connecting the bulkhead fitting assembly to the vacuum pump.
6.
Remove the Case Front and lay along side the Case Bottom.
7.
Disconnect the Wiring Harness from the PC Board
8.
Disconnect the wires from the IEC Connector
9.
If the unit contains a Docking Station Charger disconnect the wire from the Docking Station to the Power
Supply.
10. Reassemble in reverse order.