Allied Healthcare ResusciTIMER Manuel - Sayfa 3

Tıbbi Ekipman Allied Healthcare ResusciTIMER için çevrimiçi göz atın veya pdf Manuel indirin. Allied Healthcare ResusciTIMER 7 sayfaları.

Peak Airway Pressure, this number is the peak pressure delivered during each breath. It is continuously updated during the delivery
of a breath and is cleared at the start of each breath. The Peak Airway Pressure is measured in centimeters of water (cm-H
peak airway pressure display will flash "61" if pressures over 60 cm H2O are delivered to the patient.
WARNING: This device has been designed as a guide to help achieve resuscitation per the American Heart Association
Guidelines 2005. Physician's instructions or department protocols should be followed if they differ from this device.
Installation of the Resuscitation Timer with a Life Support Products L770 BMR
Warning: Do not attach the ResusciTimer to an ET tube or other airway device. The weight of the
ResusciTimer may cause the airway device to move from its intended position.
Installation of ResusciTimer with a Life Support
Products, Inc. L770 BMR:
1. Install the ResusciTimer to the corrugated tubing
using the optional hose clip or the Velcro strap.
2. Install one end of the pressure tube onto the CPR
timer hose barb as shown. See Fig. 1.0
3. Install the second end of the pressure tube onto
manometer port of the L770 BMR, as shown.
See Fig. 1.0
A ResusciTimer is shown installed on a L770 BMR
in the figure 1.0 to the right
L770 BMR
Manometer Port
L770 BMR
Figure 1.1
3
O). The
2
Figure 1.0
L770-CPR-CPL
CPR-TIMER HOSE
CLIP (ACCESSORY)
Install only on L770
corrugated Tubing.
ResusciTimer
Hose Barb
Pressure Tube