Scorpio 1800 Installation Instructions
TROUBLE SHOOTING
Some basic checks are detailed below if in doubt contact service support on
0870 066 0809.
Bath Does not move
Check Power:
Check Spur Fused to ensure there is power to the unit.
Ensure the RCCB is working and has power to it. To test the RCCB push the test
button; the RCCB should interrupt power and the switch should trip. Push the
RESET button; power should be restored. If the RCCB fails to operate in this
manner there is a problem with the electric supply and the service department
should be contacted.
If mains power is interrupted the bath will only work on battery backup.
The bath will continue to operate but as the batteries become low the bath
will emit an audible beep when handset is used.
Once the warning Beep starts, lower the bath and do not use until the fault is rectified.
Once power is restored allow the batteries to recharge before operation (usually 3 hours)
If the bath fails to operate, reset the system:
Depress both 'Bath' buttons on the handset at the same time – 10 long beeps will sound
and the system will reset. It is worthwhile checking that the handest cord and jack are still
firmly located into the control box.
Bath Judders:
Usually occurs if the bath has not been fixed to the floor correctly or the bath is not level.
Bath Noisy:
Check to ensure there is nothing touching the bath while it moves. Ensure the pipe-work
and cables are free of the panels and not rubbing against the framework. Check panels are
free to move and are not touching each other.
Bath Leaking:
Look for source of leak-typically caught flexi-hose on feed or waste. This can usually be
remedied by your own maintenance contact.
Bath is uneven (one end higher than the other)
This may occur after a power failure. Use DOWN button on the handset. Lower the bath
to its lowest position. This will automatically rest the bath so that both sides rise together.
Bath Beeping
Mains power has failed to the unit- see section titled 'Bath does not move'
Bath Panels touching/noisy
If the upper and lower panels touch, marking will occur. Adjustment is possible on the
bayonet clip if required.
12