Running a Blank
A blank must be run every 12 hours. Each blank measurement remains
valid for 12 hours of continuous operation, after which the analyzer displays
BLANK needed and a new blank is required. In addition, a new blank must
be run each time the analyzer has been powered on and off again.
• If the analyzer is in standby mode, touch the measure icon.
• Touch the Run button.
• If a blank is necessary, the "Blank needed" screen appears.
• Touch OK to confirm and run the blank.
• Check the results to see whether the blank falls within the acceptable
ranges shown in the following table. If so, press Accept.
Acceptable blanks will have
the following readouts:
WBC 0.0-0.5
RBC 0.0-0.05
PLT 0-26
HGB 0-10