Cleaning, Storage, Service
CLEANING
Clean the mattress provided with the bed by wiping down with a cloth or sponge using mild soap and
water only.
The hand pendant, scale display (optional), siderails, and bed frame can also be cleaned with a damp
cloth or sponge and warm, soapy water.
Do not use any scouring or abrasive cleaners on the bed.
CAUTION: Do not hose down, high pressure spray, or place the bed in a shower as this may cause
damage to the electronics system and void the warranty.
If any bed electrical components should become wet, disconnect bed from the 120VAC wall outlet until
the device is thoroughly dry.
STORAGE OF THE BED
This equipment is provided with a battery back up. To insure the longevity of the battery back up we
recommend that the bed be plugged into an appropriate wall outlet to keep battery charged when not in
use for extended periods of time. Failure to do so could render the battery backup system inoperable.
CAUTION: Store in a clean dry area. Dampness may affect electronic components and cause rust.
SERVICE OF THE BED
WARNING: If the bed looses any function or is damaged in any way stop using the bed
immediately and refer it to a qualified service personel. Further equipment damage or personal
injury may result.
Refer to the service chart provided in this manual for basic problem solutions. Major service of this equip-
ment should be referred to qualified personnel.
Refer to circuit layout diagram and part identification diagram provided in this manual. The manufacturer
of this equipment will provide assistance to the user's appropriately qualified technical personnel to
repair those parts of the equipment which are designated by the manufacturer as repairable. Use only
replacement components which are approved and designated as acceptable by the manufacturer.
INSPECTION
The recommended frequency of inspection for the bed is once per month or before each new patient,
whichever occurs first.
FUSES
The control box located under the left side of the head gatch has two line fuses located behind a round
removable plug located on the underside of the box. A pair of pliers can grasp the "handle" of the plug
and extract it from the box. The fuses are designated as follows:
5x20mm 1.6L250V Time-Lag fuse (1.6A/250V) Wickmann Series 19195 or equivalent
16 TRI-FLEX II BED - OPERATION MANUAL