• Detach the long sterilisable silicone tubes Ø 8 x 14 - long (5.) from the vessel (4.). Disconnect the sterilisable silicone tubes Ø 8
x 14 - short (3.) from the respective vessels and from the respective antibacterial filters and empty the vessel/s. The liquids col-
lected in the vessels must be disposed of in the biological waste container.
• In order to clean the accessories, rinsing them under hot running water is sufficient. Before performing the vessel cleaning opera-
tions empty their content into the biological material vessels and clean them by sterilisation in an autoclave with a cycle at 121°C.
The silicone aspiration tubes and aspiration fitting can be sterilised in an autoclave with a cycle at 121°C. The complete vessel
should be replaced every 30 sterilisation cycles.During the sterilisation operations the operator must use latex gloves and suitable
clothing in order not to come into contact with any possibly contaminating substances.
• The disposable antibacteria filter (2.) must be replaced after every application.
• Never wash the device under water or by immersion; clean the outside of the device using a cloth moistened with detergent (non
abrasive).
MAXIASPEED does not require maintenance and/or lubrication. Nevertheless, several simple several simple checks must be performed
before every use:
• Check the integrity of the casing, power supply cable and the pedal switch, if supplied.
• Block the aspiration connector with a finger and check that the vacuum level reaches -0.9 bar.
• Check that there are no disturbing noises suspicious of malfunctioning.
• Check that the cage is positioned correctly inside its housing. It must be in axis with the plug aspiration hole so that the floating valve
can block it in the circumstance in which the aspirated liquid exceeds the maximum volume the vessel can receive. Do not connect
the cage as shown in figure 3b; it must be connected as shown in figure 3a.
• Check that the float is fitted in the correct position and that it is free to move inside the cage (dirt or encrustation may hinder its
movement). Figure 4 shows the correct insertion.
• Check that the plug seal is inserted with the edging as illustrated in figure 5.
PROBLEMS
Excessive noise
The unit turns on but it does not suck
The vacuum value cannot be adjusted
The protection fuse is activated any time
the device is turned on
The vacuum gauge does not work
Note: if you experience faults or malfunctioning problems different to those listed above, always and exclusively
contact authorised assistance centres.
CLEANING AND DISINFECTION OPERATIONS
PERIODICAL CHECK FOR THE SAFETY OF THE DEVICE

PROBLEMS, CAUSES and SOLUTIONS

Damaged pump or obstructions in the
internal suction duct
- Damaged pump
- Vacuum regulator fully open.
Connection tubes disconnected
and/or badly connected, broken
connection tubes. Vessel not in a
vertical position, full, or defective
overflow valve. Possible blockage of
the hydraulic circuit inside the unit
Damage to the internal hydraulic circuit or
obstruction of the connection tubes to the
aspiration unit
Pump damaged or in short circuit
Liquid penetrating the pneumatic
circuit
POSSIBLE CAUSES
11
SOLUTIONS
Send to the customer care service
- Send to the customer care service.
- Check the position of the vacuum
regulator. Check the connections and the
integrity of the tubes.
Position the vessel in a vertical position,
check the overflow valve (blocked) and/or
replace the silicon tubes
Send to the customer care service
Send to the customer care service
Send to the customer care service