3A HEALTH CARE MYNEB Kullanım Kılavuzu - Sayfa 5
Solunum Ürünleri 3A HEALTH CARE MYNEB için çevrimiçi göz atın veya pdf Kullanım Kılavuzu indirin. 3A HEALTH CARE MYNEB 8 sayfaları.
Disinfecting
All nebuliser components (except air hose) can be disinfected with chemical disinfectants following dosage and use
limitations as provided by disinfectant manufacturer. Disinfectants are usually available at pharmacies.
sterilizing with steam
All nebuliser components (except air hose and masks) can be heat steam sterilized up to 121 °C (20 min.) or 134 °C
(7 min.). EN554/ISO11134.The sterilization packaging must conform to EN868/ISO11607 and be suitable for steam
sterilization. After sterilization always let all components cool down to ambient temperature before further use.
Do not repeat sterilization cycle when components are still warm.
MAINTENANCE, CARE, AND sERVICE
Order all spare parts from your dealer or pharmacist.
- It is advisable replace nebulizer after some 100 to 120 treatments on single patient or after about 20 sterilization
cycles. Check the filter continually for cleanliness and replace it if dirty, or after a maximum of 3 months use. Spare
filters are provided with the device.
- To replace the air filter (1b), extract the air filter compartment (1a) from the device using a screwdriver (Fig. F).
Exchange the used air filter with a new filter. Gently place the new filter into the inner part of the compartment (Fig. G)
and make sure that it is correctly fixed.
MAlFUNCTIONs AND ACTIONs TO TAKE
• The device cannot be switched on
Ensure the AC adapter (8) is correctly plugged to the AC adapter socket (1e) of the device and the power plug into the
wall socket. Ensure the ON/OFF switch (1d) is in the position «I». (Fig. C)
• The nebuliser functions poorly or not at all
Ensure the air hose (2) is correctly connected at both ends.
Ensure the air hose is not squashed, bent, dirty or blocked. If necessary, replace with a new one.
Ensure the nebuliser (3) is fully assembled and the vaporiser head (3a) is placed correctly. (Fig. B).
Ensure the required medication has been added. (Fig. D)
GUARANTEE
This device is covered by a 3 year guarantee from the date of purchase. The guarantee is valid only on presentation
of the guarantee card completed by the dealer (see back) conforming date of purchase or the receipt. The guarantee
covers only the compressor. The replaceable components like nebuliser, masks, mouthpiece, air hose, and filters are
not included. Opening or altering the device invalidates the guarantee. The guarantee does not cover damage caused
by improper handling, accidents or noncompliance with the operating instructions.
IMPORTANT INFORMATION REGARDING ElECTRO MAGNETIC COMPATIBIlITY (EMC)
Electronic devices (PC's, mobile phones, televisions, etc) that generate electromagnetic interference can affect the
correct functioning of medical devices. This situation can lead to incorrect functioning creating unsafe conditions.
In order to show compliance with the Directives on Electromagnetic Compatibility (EMC), all the tests introduced by
the standard EN60601-1-2:2007 have been implemented. This standard defines the levels of immunity to electromag-
netic interference as well as maximum levels of electromagnetic emissions for medical devices: 3A HEAlTH CARE
products comply with the levels specified in the standard.
However, when the device is in use, it is best practise not to use mobile phones or other electronic devices which
generate strong electromagnetic fields near the medical device. This may create dangerous conditions that put the
successful outcome of treatment at risk. On the basis of experience and tests conducted, it is recommended that
a minimum distance of 7 m be kept: if the distance is less, the correct functioning of the device should be verified.
Documentation regarding the standard EN60601-1-2:2007 is available from 3A HEAlTH CARE (see the address in the
instruction manual) and on the website www.3-a.it.
9