8. Connect the nasal prongs to the child as below:
9. Dial the flows of air and oxygen required to see a constant stream of bubbles in the water.
Start with 5 L/min of oxygen and 5 L/min of air. Check and record the flows dialled and the
oxygen concentration this provides. A total flow of 10 L/min is sufficient in most cases to
deliver CPAP, which you can see by continuous bubbling in the bottle.
10. Check the child for signs of respiratory distress, check the SpO
in the bottle.
a. If the SpO
the CPAP level to 8 or 10cmH20 and then, if no response the oxygen flow meter to
as much as 8 L/min.
b. If there are not continuous bubbles, check the nasal prongs are attached properly,
and reposition them so they fit snugly inside the nostrils.
c. If there are still not continuous bubbles, check for leaks along the circuit, and adjust
the oxygen or air flows according to the chart below.
a) Place the bonnet on
patients head, centering
the "securement line
indicator"
b) With the bonnet
centered at nape of neck,
place remaining bonnet
on infant
c) Close the top of bonnet by
twisting bonnet until
tightly secure to head
crown.
d) Secure with ribbon
provided
e) Cut self-adhesive Velcro in
half and secure to the
bonnet
is below 90% or the child has severe respiratory distress, first increase
2
© Copyright Diamedica (UK) Ltd 2018
f) Attach chin strap to
the infant and secure
to Velcro
g) Place the nasal
prong into the
infants nostril
h) Secure tubing by
placing elastic strap
over tubing and secure
to Velcro. Elastic strap
is correctly positioned
when strap is in line
within the securement
line indicator
, and check if there are bubbles
2
Page 5 of 13