ABB EZClean Kullanım Talimatları Kılavuzu - Sayfa 12
Endüstriyel Ekipmanlar ABB EZClean için çevrimiçi göz atın veya pdf Kullanım Talimatları Kılavuzu indirin. ABB EZClean 16 sayfaları. Compressed air supply unit
12
E Z C L E A N | C O M P R E S S E D A I R S U P P LY U N I T | O I/ E ZC L E A N - E N R E V. A
...5 Configuration
Sensor Setup
Menu
Sensor Setup
Exit
Select
Menu
S1 (to 4) :DO
Tag
PV Type
Units
Range High
Range Low
Filter Type
Clean Interval
Clean Type *
Clean On Time *
Clean Off Time * / **
Number of Pulses * / **
Recovery Time *
Clean Duration *
Clean Output *
Salinity Correction
Barometric Pressure
Restore Defaults
Salinity Units
Pressure Units
* Displayed only if Clean Interval is NOT set to Off
** Displayed only if Clean Type is set to Pulsed
Used to set the tag, measurement units, operational range and clean functions and to
compensate for salinity and barometric pressure.
Comment
Select the ID of the optical dissolved oxygen probe to be cleaned and scroll to the
Clean Interval menu to configure the cleaning routine.
These parameters are not used to configure cleaning – refer to the probe Operating
instruction (OI/ADS420-EN) for parameter descriptions.
Set the interval between cleans:
Off / 15 Mins. / 30 Mins. / 45 Mins. / 1 to 24 Hours
Set the clean type: Continuous / Pulsed
Set the duration of the clean: 1 to 60 Secs
Set the duration between cleans: 1 to 60 Secs
Set the number of cleaning pulses: 1 to 10 Pulses
Set the time delay between the completion of cleaning and the display of a new
reading on the operator page: 1 to 10 Min.
Displays the total duration of the clean:
Clean Type set to Continuous = Clean on Time + Recovery Time
Clean Type set to Pulsed = (Clean on Time + Clean Off Time)
x Number of Pulses + Recovery Time
Displays the output signal the clean is assigned to. This can be set to relay 1 to 6 or
digital output 1 to 6 – refer to AWT420 transmitter Operating instruction
(OI/AWT420-EN).
These parameters are not used to configure cleaning – refer to the probe Operating
instruction (OI/ADS420-EN) for parameter descriptions.
Default
Off
Continuous
30 Secs
30 Secs
2 Pulses
1 Min.
No Assignment