Guidance and manufacturers declaration - electromagnetic immunity
The Biomet
SpinalPak
®
electromagnetic environment specified below. The customer or the user of the Biomet
Non-invasive Spine Fusion Stimulator System should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 610004-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
( 50/60 Hz )
magnetic field
IEC 61000-4-8
Non-invasive Spine Fusion Stimulator System is intended for use in the
®
IEC 60601
Compliance level
test level
± 6 kV contact
± 8 kV air
Not Applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Table 2
Floors should be wood, concrete or
ceramic tile. If floors are covered with
Group 1
synthetic material, the relative humidity
Class B
13
Electromagnetic environment -
guidance
should be at least 30 %.
SpinalPak
®
®