OPERATOR'S MANUAL
NOTE: Refer to the Normal
Operation section for
additional information on
the screen displays of
Selecting and Testing
Grain.
®
mini GAC
11001-1464A-201701 Rev B
QUICK START GUIDE
IMPORTANT:
The cell must be empty and the loading cup
removed before turning the unit ON and
during all testing (progress bar indicates
when testing).
SELECTING GRAIN
1. Press the On/Off/Home button to power on the unit.
2. To select a grain type, press the Down Arrow button to
position cursor at the Product heading and press Enter.
3. Select a product (grain) using the Up or Down Arrow
button and press Enter.
4. The selected grain will appear at the top of the display.
GRAIN SAMPLE TESTING
1. Ensure Loader is removed from the top of the unit.
2. With the product name selected at the top of the display,
press the Enter button.
3. The cell must be empty, upright, and still to perform an
empty cell test. When the Empty Cell menu displays, press
the Enter button.
IMPORTANT:
The empty cell reference measurement is
used for all product testing until the unit is
powered off. It is critical that the cell be
clean and completely empty when the unit is
on. Keep the measurement cell opening
clear of hands or other objects during this
period.
4. After the empty cell test completes, the Fill Cell menu
displays.
5. With the slide closed, scoop or pour grain into the Loader
over the minimum fill line (Label on loader identifies
minimum fill line)
6. Place the Loader on top of the unit.
7. When secured, pull Loader slide out to dispense grain into
the unit.
8. Remove the Loader and use the level edge to strike
any excess grain from the unit.
9. Press Enter to begin the test.
(Figure
5).
QUICK START GUIDE / 5