The IR temperature sensor can be cleaned with the Q-tip by placing the
Q-tip through the funnel or on the outside of the funnel as depicted in
(Figure
22).
Figure 22
Cleaning the Sensor with Cotton Swab
Funnel
6. Allow the temperature sensor to dry for 90 seconds then gently clean
with the dry end of the cotton swab.
7. Visually inspect the temperature sensor as well as all other areas
cleaned to ensure the instrument is free from debris. If there is any
question about the cleanliness of the instrument, it should be returned
to a dealer or authorized service center.
8. If cleaning is acceptable, return instrument to upright position and
replace the drawer.
9. Reconnect power cable and accessory cords.
10. Power on instrument.
IMPORTANT: The foregoing recommendations are provided as a
guideline to maintain a robust and quality operating
GAC 2500. It should not be interpreted as an exhaustive
maintenance program. Dust and debris may periodically
accumulate in areas not specified in this manual. The
owner is responsible for ensuring overall equipment
cleanliness. If any questions arise regarding the
maintenance or performance of the instrument, contact
your dealer or local authorized service center.