Dickson Industrial SC3 Kullanıcı Kılavuzu - Sayfa 3
Kayıt Ekipmanları Dickson Industrial SC3 için çevrimiçi göz atın veya pdf Kullanıcı Kılavuzu indirin. Dickson Industrial SC3 8 sayfaları. Temperature chart recorder
Product Specifications
Temperature Sensor:
Recorder Accuracy:
Ambient Operating Temperature: -22 to +122˚F (-30 to +50˚C)
Response Time:
Temperature Ranges:
Recording Time:
Power Supply:
Resolution:
Mounting:
Chart Size:
Dimensions:
Weight:
Includes:
Rev. 06/03
Precise spiral wound bi-metallic strip transducer
±4˚F (±2.5˚C) over full scale
11 minutes to move 63% of full scale in air moving 5' per second
(˚F or ˚C depending on chart installed) -14 to +32˚F (-25 to 0˚C),
+4 to +50˚F (-15 to +10˚C), +22 to +68˚F (-5 to +20˚C),
+50 to +96˚F (+10 to +35˚C), +76 to +122˚F (+25 to +50˚C)
24 hour or 7 day
Single "AA" battery, up to 2 years average life
Not less than 1" per +55˚F in 1˚C or 2˚F steps
Free standing, wall mounted or suspended
3" (75mm) DIA
3.667" x 3.667" x 2.232" (93mm x 93mm x 57mm)
0.5 lbs. (227 g)
Battery, pen and instruction manual