3. Instruction for use

3-4. Stimulation with the MED64-Allegro System

The MED64-Allegro System allows you to stimulate each well sequentially. Use Mobius Acquire MED64R2
Data w/Stim's Step1-8 to stimulate one (or 2 channels) per well (sample) sequentially. Inactivate the
Step 5-8 when the 4 Well Probe is used.
Set the Trace interval (stimulus interval) for longer than 5 seconds since the amplifiers need at
least 5 seconds to recover after stimulation.
Stim
Stim
Stim
Stim
Stim
Figure 13. An example for stimulation to the 4 Well Probe. Stimulation is applied to Well#1-4 with interval of 10
seconds.
Product Manual - MED Multi-well Connector (MED-C11)