6. Cautions

5-2. Sterilization

Wipe the Multi-well Connector with a lint-free cloth soaked in 70% ethanol, and allow to dry.
CAUTION:
Do NOT autoclave as it may damage the MED Multi-well Connector.

5-3. Storage

Store in a cool dry area. Avoid exposure to high temperatures or humidity.
6. Cautions
DO NOT spill medium or any other liquid on the contact pins.
DO NOT give strong mechanical shock by putting heavy material on the unit or dropping the unit.
DO NOT touch the contact pins on the connector unit with bare hands or fingertips when they are
cleaned. This may cause rust and adversely affect conduction.
What should I do if medium is spilled onto the contact pins?
Stop your experiment immediately. Disassemble your Connector and clean the contact
pins with kim-wipe soaked in 70% ethanol, and allow to dry.